    AgendaAgenda

    Mounting the AutoRAE Lite

    Powering the AutoRAE Lite

    Connecting calibration gas

    SD Card

    Bump testing & calibration

    Transferring data AutoRAE II Utility program

    Updating AutoRAE Lite Firmware

  • 8/3/2019 Kurzanleitung ARL2 En

    3/91

    RAE Systems Protection Through Detection

    RAE Systems Protection Through Detection

    AutoRAE LiteAutoRAE Lite

    Instrument

    dock (boot)

    Bump and calibration

    activation buttons and

    function indicator LEDs

    Calibration

    gas gauge

    Gas cylinder dock

    Status LEDs

    Slide

    mechanism

  • 8/3/2019 Kurzanleitung ARL2 En

    4/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Mounting the AutoRAE LiteMounting the AutoRAE Lite

    Total clearance width:

    525 mm (21)

    Total clearance height :

    702mm (27.75)

  • 8/3/2019 Kurzanleitung ARL2 En

    5/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    AC Adapter (Power)AC Adapter (Power)

    The AutoRAE Lite can be powered by AC adapter or four C-size alkalinebatteries.

    Anytime the batteries are replaced, the AutoRAE Lite is unplugged, or the

    power has been interrupted for more than 7-10 days, the date may needto be reset using the AutoRAE Lite Utility Program.

  • 8/3/2019 Kurzanleitung ARL2 En

    6/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Alkaline BatteriesAlkaline Batteries

    1. Push tab

    2. Pull up

    3. Remove door

    4. Insert 4 C alkaline batteries (Q: Can I leave the

    batteries in the ARLite? A: Yes, it will not damage the

    station)

  • 8/3/2019 Kurzanleitung ARL2 En

    7/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Connecting Calibration GasConnecting Calibration Gas

    Place the cylinder in the gas cylinder dock and then turn it

    clockwise until it is snug

    Once the cylinder is attached, the gauge should immediately

    give a reading

    The reading should be equal to or greater than 50 psi

    (approximately 345 kPa or 3.45 bar). If the reading is lower or shows zero, then the tank does

    not have sufficient contents and must be replaced.

  • 8/3/2019 Kurzanleitung ARL2 En

    8/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Connecting Calibration GasConnecting Calibration Gas

    1. Insert calibration

    gas cylinder

    2. Screw-tighten

    cylinder in place

    3. Cylinder in

    place

  • 8/3/2019 Kurzanleitung ARL2 En

    9/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Connecting Calibration GasConnecting Calibration Gas

    Note: The AutoRAE Lite is designed for 58L

    and 108L cylinders of calibration gas.

    It also works with a 34L cylinder, but anadapter sleeve is needed in order to keep

    the cylinder securely docked in the AutoRAE

    Lite.

    Part number: T01-2051-000

    Price: ???

  • 8/3/2019 Kurzanleitung ARL2 En

    10/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Connecting Calibration GasConnecting Calibration Gas

    For Fresh air calibration, use a filter on

    the AutoRAE Lite inlet

    The inlet is located on the left of theUSB and SD card ports

    Replace filter on a regular basis or if it

    is discolored

  • 8/3/2019 Kurzanleitung ARL2 En

    11/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    SD Card (make it click)SD Card (make it click)

    IMPORTANT!

    The AutoRAE Lite cannot operate without an SD card in the port.

    SD port cover closed.

    SD port cover opened.

    Press SD card into

    port until it clicks

  • 8/3/2019 Kurzanleitung ARL2 En

    12/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    SD CardSD Card

    When SD card is

    correctly in the port,

    the: Service Request,

    LED, Bump, and Cal

    LEDs glow green.

    If no SD card is in port,

    the AutoRAE Lites

    Service Request LED

    glows red.

  • 8/3/2019 Kurzanleitung ARL2 En

    13/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    SD CardSD Card

    If SD card is locked,

    LED glows orange and

    blinks.

    The AutoRAE Lite

    cannot write data to a

    locked SD card.

    Remove SD card and

    unlock by moving lock

    tab; then reinsert SD

    card.

    Locked

    Unlocked

  • 8/3/2019 Kurzanleitung ARL2 En

    14/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    TestingTesting

    Turn on the QRAE II by pressing

    and holding the [MODE] key

    Once the QRAE II is fully turnedon and is reading

    Verify FIRMWARE and if QRAE II

    is pump or diffusion (make

    appropriate changes, if needed,to the AutoRAE Lite).

  • 8/3/2019 Kurzanleitung ARL2 En

    15/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    FirmwareFirmware

    Current firmware: v.2.05

    Upgrade firmware: v3.00

    Users must upgrade QRAE II to work with AutoRAE Lite

    Subject: AutoRAE Lite QRAEII release

    FAQ:

    Q: What is the most current firmware for the QRAEII?

    A: v2.05

    Q: When did we first ship firmware v2.05?

    A: January 2009

  • 8/3/2019 Kurzanleitung ARL2 En

    16/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Q: Will firmware v2.05 allow the QRAEII to work with the AutoRAELite?

    A: No, you must upgrade to firmware v3.00

    Q: When will firmware v3.00 be available?

    A: 25 May 2009

    Q: Where do I get a copy of firmware v3.00?

    A: Download page ofwww.raesystesm.com

    Q: I have a QRAEII and I want to use it in the AutoRAE Lite, can I

    upgrade it? A: Depends on serial number

  • 8/3/2019 Kurzanleitung ARL2 En

    17/91

  • 8/3/2019 Kurzanleitung ARL2 En

    18/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Q: What if my front pump plate does not have the filter stem orexternal filter?

    A: A conversation kit can be purchased (part number: XXX-XXXXX-

    XXXX)

    Q: If I have an older generation QRAEII, will it work with the AutoRAELite?

    A: Diffusion models with 155-100001 or before AND/OR Pump models

    with 181-000001 or before will not be available for upgrade.

    OLD NEW

  • 8/3/2019 Kurzanleitung ARL2 En

    19/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Q: Can I download the firmware to my QRAEII using the AutoRAE

    Lite?

    A: No, you will need to use the ProRAE Studio

    Diffusion units: QRAEIIs within a certain serial number range can

    be upgraded and will work with the AutoRAE Lite, if a diffusion

    adapter is used. Pump units: QRAEIIs within a certain serial number range can be

    upgraded (and depending on the front pump piece) and will work

    with the AutoRAE Lite.

  • 8/3/2019 Kurzanleitung ARL2 En

    20/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Inserting the diffusion adapterInserting the diffusion adapter

  • 8/3/2019 Kurzanleitung ARL2 En

    21/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Diffusion setting or Diffusion setting or

  • 8/3/2019 Kurzanleitung ARL2 En

    22/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    .Pump (this is the default setting).Pump (this is the default setting)

  • 8/3/2019 Kurzanleitung ARL2 En

    23/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Turn ON and screen DOWNTurn ON and screen DOWN

  • 8/3/2019 Kurzanleitung ARL2 En

    24/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Turn ON and screen DOWNTurn ON and screen DOWN

  • 8/3/2019 Kurzanleitung ARL2 En

    25/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    QRAE II (screen down)QRAE II (screen down)

  • 8/3/2019 Kurzanleitung ARL2 En

    26/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Bump and/or CalibrateBump and/or Calibrate

    Firmly press the QRAE II facedown into the dock on the AutoRAE Lite.

    The two LEDs above the buttons should glow green. If not, make sure the

    QRAE II is firmly seated.

  • 8/3/2019 Kurzanleitung ARL2 En

    27/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Slide in OPEN and CLOSED positionSlide in OPEN and CLOSED position

  • 8/3/2019 Kurzanleitung ARL2 En

    28/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Bump or CalibrationBump or Calibration

    Press Bump to perform a bump test.

    Press Cal. to perform a calibration.

    You will be alerted when the bump test or calibration is complete.

    When the two indicator LEDs above the buttons glow green, remove the

    QRAE II from the dock.

  • 8/3/2019 Kurzanleitung ARL2 En

    29/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Bump or CalibrationBump or Calibration

    Fail: Red LED indicates

    bump test or calibration

    failed.

    If bump test or

    calibration fails, check

    gas pressure and tryagain.

    If it fails again, it is

    possible the sensor must

    be replaced.

  • 8/3/2019 Kurzanleitung ARL2 En

    30/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Bump or CalibrationBump or Calibration

    Pass: Green LED

    indicates bump test or

    calibration wassuccessful.

  • 8/3/2019 Kurzanleitung ARL2 En

    31/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Last Step -Last Step -

    After a successful BUMP or

    CAL, open the slide,remove* the QRAE II, and

    press either button to return toNormal Mode (sensing gas).

    *Wait for 1-3 seconds toremove or audible will sound

  • 8/3/2019 Kurzanleitung ARL2 En

    32/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Transferring DataTransferring Data

    Calibration and event data is

    collected each time the QRAE II

    is bump tested or calibrated.

    The AutoRAE Lite stores data

    on the SD card, providing

    convenient storage and easy

    data transfer.

    Some computers will read the

    SD card directly, others will

    need an adapter to plug into

    USB port.

  • 8/3/2019 Kurzanleitung ARL2 En

    33/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Transferring DataTransferring Data

    Data stored on the SD card is in standard ASCII textformat, making it easy to import into most spreadsheetand word-processing programs

    Data is in CSV (comma-separated value) form so that itcan be put in existing spreadsheets (Excel) or intocolumns in word processing programs

  • 8/3/2019 Kurzanleitung ARL2 En

    34/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Removing SD CardRemoving SD Card

    Pry up the corner of the

    protective cover and open

    Press on the SD card to release

    Remove the card from the

    AutoRAE

    Insert into computer or an SD

    card adapter and then plug intoUSB port

  • 8/3/2019 Kurzanleitung ARL2 En

    35/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Removing SD CardRemoving SD Card

    Do not remove the SD

    card from the AutoRAE

    Lite during a bump testor calibration.

    After the two LEDs

    above the buttonsglow green, it is safe to

    remove the SD card.

  • 8/3/2019 Kurzanleitung ARL2 En

    36/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    AutoRAE Utility ProgramAutoRAE Utility Program

    Read the SD Card

    Check and configure the AutoRAE Lite

    With the QRAE II installed in the AutoRAE Lite, configure

    the QRAE II

    (some features may use the Utility Program and others

    may use the ProRAE Studio)

  • 8/3/2019 Kurzanleitung ARL2 En

    37/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Reading the SD CardReading the SD Card

    Once the SD card is in the adapter and connected to thecomputer, it should show up as an external device or adrive.

    Double-click the My Computer icon on your desktop.

    When it opens, you should see a list of drives, includingyour C:, where most programs reside, and the external

    device holding the SD card.

  • 8/3/2019 Kurzanleitung ARL2 En

    38/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Reading the SD CardReading the SD Card

  • 8/3/2019 Kurzanleitung ARL2 En

    39/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Reading the SD CardReading the SD Card

    Double-click on the name of the device to open its

    directory. You now see three folders containing files

    created each time the AutoRAE Lite was turned off or theSD card was inserted.

    Create a new folder on your desktop or elsewhere on your

    computer.

    Drag the file(s) from the SD card into the folder on your

    computer:

  • 8/3/2019 Kurzanleitung ARL2 En

    40/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Reading the SD CardReading the SD Card

  • 8/3/2019 Kurzanleitung ARL2 En

    41/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Reading the SD CardReading the SD Card

    Open the folder to make sure the files have transferred

    correctly.

    Inside each folder is a file with an 8-character name thatstarts with the letter of the type of file (bump test,

    calibration data, or events).

    Note: Once the files are transferred from the SD cardto your computer, you may change the name of the file

    to suit your needs.

  • 8/3/2019 Kurzanleitung ARL2 En

    42/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Reading the SD CardReading the SD Card

    The first two digits indicate the year, the second two digits

    indicate the month, and the ending of _V1 is a reserved tag.

    For example, an events file appears like this: E0802_V1 This indicates that it is an event log from the year 2008, and

    the month February.

  • 8/3/2019 Kurzanleitung ARL2 En

    43/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    AutoRAE Utility ProgramAutoRAE Utility Program

    When connected to a computer with a USB cable, the

    AutoRAE Lites parameters can be viewed

    In Administrator Mode, parameters can be changed These include synchronizing the AutoRAE Lites time to

    match the computers clock.

  • 8/3/2019 Kurzanleitung ARL2 En

    44/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    AutoRAE Utility ProgramAutoRAE Utility Program

    Install the USB driver.

    Follow instructions on the RAE Systems website or on

    utilities disc that accompanies AutoRAE Lite. Install the AutoRAE Lite Utility Program on your computer.

    Follow instructions after downloading the program from

    RAE Systems website or after you insert the disc intoyour computer.

    Once installation is complete, there should be an icon for

    the Utility program on the desktop:

  • 8/3/2019 Kurzanleitung ARL2 En

    45/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    AutoRAE Utility ProgramAutoRAE Utility Program

    Connect the AutoRAE Lite to a computer using a USB cable.

    Configure your computers port.

    Follow this path: Start >> Settings >> Control Panel >>System.

    You now see the System Properties dialog box.

    Click Hardware.

  • 8/3/2019 Kurzanleitung ARL2 En

    46/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    AutoRAE Utility ProgramAutoRAE Utility Program

  • 8/3/2019 Kurzanleitung ARL2 En

    47/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    AutoRAE Utility ProgramAutoRAE Utility Program

    Click Device Manager.

    You should see a list of devices in and attached to your

    computer. Scroll down the list until you see Ports.

    Open Ports (click on the + to the left of it) and you should

    see the port for the AutoRAE Lite, plus its port number (forexample, COM6).

    Write down this port number; you will need it when you start

    AutoRAE Utility Program.

  • 8/3/2019 Kurzanleitung ARL2 En

    48/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    AutoRAE Utility ProgramAutoRAE Utility Program

  • 8/3/2019 Kurzanleitung ARL2 En

    49/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    AutoRAE Utility ProgramAutoRAE Utility Program

    Close Device Manager and System Properties.

    Double-click the AutoRAE Utility Program icon on your

    desktop to start the program.

  • 8/3/2019 Kurzanleitung ARL2 En

    50/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    AutoRAE Utility ProgramAutoRAE Utility Program

  • 8/3/2019 Kurzanleitung ARL2 En

    51/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    AutoRAE Utility ProgramAutoRAE Utility Program

    Pull down Operation and select

    Setup Port.

    Click the COM(5-256) button with

    your mouse.

    Select COM port setting found when

    you went into the computers System

    Properties.

    Click OK.

  • 8/3/2019 Kurzanleitung ARL2 En

    52/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    AutoRAE Utility ProgramAutoRAE Utility Program

    The programs screen consists of a menu bar, toolbar, and aleft selection/navigation pane, and two right informationpanes for synopsis and information.

    In addition, when the program is in Administrator Mode, thelower right pane becomes an interface for setting parametersin a QRAE II.

    At the bottom of the screen is a Status Bar, which tells whenthe program is ready.

  • 8/3/2019 Kurzanleitung ARL2 En

    53/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    AutoRAE Utility ProgramAutoRAE Utility Program

    Menu bar

    Tool bar

    Synopsis of

    settings

    Selection/

    Navigation

    Information /

    Parameter Setting

    Status bar

  • 8/3/2019 Kurzanleitung ARL2 En

    54/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Menu BarMenu Bar

    The menu barconsists of five pull-down

    menus:

    File

    Operation

    Option

    View

    Help

  • 8/3/2019 Kurzanleitung ARL2 En

    55/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Menu BarMenu Bar

    File. The File menu includes:

    Save Save the current file

    Print Print the synopsis information Print Preview View the page to be printed before printing

    Print Setup Standard choices for setting the pagesparameters for printing

    Exit Exit the AutoRAE Utility Program

  • 8/3/2019 Kurzanleitung ARL2 En

    56/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Menu BarMenu Bar

    Operation. These are functions for communication andcontrolling the AutoRAE Lites and ToxiRAE 3s parameters.

    Setup Port Set the COM (communication)port

    Get Configurations Open saved configurations inthe program window

    Operation Mode Go to User Mode or AdministratorMode

    Load Monitor Config Open a ToxiRAE 3 configurationfile

    Upgrade Firmware Upgrade the AutoRAE Litesfirmware

  • 8/3/2019 Kurzanleitung ARL2 En

    57/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Menu BarMenu Bar

    Option. This provides a submenu that includes language,

    which allows you to choose a language in which to operate.

    Language Select a language for the program.

    M B

  • 8/3/2019 Kurzanleitung ARL2 En

    58/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Menu BarMenu Bar

    View. Make the Tool Bar and Status Bar viewable or hidden.

    Toolbar

    Status Bar

    M BM B

  • 8/3/2019 Kurzanleitung ARL2 En

    59/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Menu BarMenu Bar

    Help. This is information to aid in using the AutoRAE Lite.

    About AutoRAE II Utility Program Displays programsversion number

    Help Window pops up with assistance.

    T lbToolbar

  • 8/3/2019 Kurzanleitung ARL2 En

    60/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    ToolbarToolbar

    The toolbarconsists of six clickable

    icons, from left to right:

    RECV (Get Configurations)

    Save The Active Document

    Print The Active Document

    Load Monitor Configuration

    (Administrator Mode only) Upgrade Firmware

    Display Program Information

    T lbToolbar

  • 8/3/2019 Kurzanleitung ARL2 En

    61/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    ToolbarToolbar

    These icons are duplicates of major functions in the menu

    bar

    If the menu bar is turned off, these provide a quick,convenient way to activate major functions.

    Note that the Configuration icon is not functional in User

    Mode and is therefore grayed out.

    AutoRAE Utility ProgramAutoRAE Utility Program

  • 8/3/2019 Kurzanleitung ARL2 En

    62/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    AutoRAE Utility ProgramAutoRAE Utility Program

    The AutoRAE Utility Program operates in two modes, User

    Mode and Administrator Mode.

    In both modes, it can be used to check the AutoRAE Lite orthe AutoRAE Lite and a QRAE II monitor.

    In Administrator Mode, settings for the AutoRAE Lite and

    QRAE II can be set.

    AutoRAE Utility ProgramAutoRAE Utility Program

  • 8/3/2019 Kurzanleitung ARL2 En

    63/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    AutoRAE Utility ProgramAutoRAE Utility Program

    With the main window open, pull down

    Operation and select Get Configurations.

    As the connection is being made and the

    configuration data is transferred, a pop-upwindow with a progress bar is shown.

    Once the information is acquired, the

    main screen is populated with information:

    AutoRAE Utility ProgramAutoRAE Utility Program

  • 8/3/2019 Kurzanleitung ARL2 En

    64/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    AutoRAE Utility ProgramAutoRAE Utility Program

    AutoRAE Utility ProgramAutoRAE Utility Program

  • 8/3/2019 Kurzanleitung ARL2 En

    65/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    AutoRAE Utility ProgramAutoRAE Utility Program

    The main screen is populated withdata and now includes icons of theQRAE II and its sensor, along withthe QRAE IIs serial number

    The other panes are populatedwith data about the AutoRAE Lite

    AutoRAE Utility ProgramAutoRAE Utility Program

  • 8/3/2019 Kurzanleitung ARL2 En

    66/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    AutoRAE Utility ProgramAutoRAE Utility Program

    AutoRAE Utility ProgramAutoRAE Utility Program

  • 8/3/2019 Kurzanleitung ARL2 En

    67/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    AutoRAE Utility ProgramAutoRAE Utility Program

    AutoRAE Utility ProgramAutoRAE Utility Program

  • 8/3/2019 Kurzanleitung ARL2 En

    68/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    AutoRAE Utility ProgramAutoRAE Utility Program

    Administrator ModeAdministrator Mode

  • 8/3/2019 Kurzanleitung ARL2 En

    69/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Administrator Mode

    Administrator ModeAdministrator Mode

  • 8/3/2019 Kurzanleitung ARL2 En

    70/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    When the pop-up window appears,

    you must type in a password.

    Note: The default password is rae.

    Click OK to register your

    password and enter Administrator

    Mode.

    Administrator ModeAdministrator Mode

  • 8/3/2019 Kurzanleitung ARL2 En

    71/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Once your password is accepted, the pop-up dialog box

    disappears and the main screen is shown.

    Click Get Configurations. As the connection is being made and the AutoRAE Lite

    configuration data is transferred, a succession of three

    animated progress pop-up windows is shown.

    Administrator ModeAdministrator Mode

  • 8/3/2019 Kurzanleitung ARL2 En

    72/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Gas Lot No

    Gas Lot number, as shown on the side of a gas cylinder

    Clock From PC Synchronize the clock in the AutoRAE Lite to the PCs time

    Cal If Bump Fails

    Tell AutoRAE Lite to perform a calibration if a bump testfails

    Administrator ModeAdministrator Mode

  • 8/3/2019 Kurzanleitung ARL2 En

    73/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Administrator ModeAdministrator Mode

  • 8/3/2019 Kurzanleitung ARL2 En

    74/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Date/Time

    Always the user to sync with computer time/date

    Password Changes the password in the QRAE II

    Administrator ModeAdministrator Mode

  • 8/3/2019 Kurzanleitung ARL2 En

    75/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Administrator ModeAdministrator Mode

  • 8/3/2019 Kurzanleitung ARL2 En

    76/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Administrator ModeAdministrator Mode

  • 8/3/2019 Kurzanleitung ARL2 En

    77/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Calibration & Bump Intervals

    Sets the number of days until the monitor requirescalibration or bump testing

    Calibration Parameters

    Sets the Calibration Span reference value

    Alarm Levels

    Sets the threshold concentrations at which the High, Low,STEL, and TWA alarms sound (refer to the QRAE II UsersGuide for further explanation)

    Administrator ModeAdministrator Mode

  • 8/3/2019 Kurzanleitung ARL2 En

    78/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Remove the QRAE II.

    Place another in the dock.

    Press Set.

    If the parameters are successfully

    transferred, you see this dialog box:

    Load Monitor ConfigurationLoad Monitor Configuration

  • 8/3/2019 Kurzanleitung ARL2 En

    79/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    You can load a monitorconfiguration from your computer toa QRAE II.

    This is useful when you want manyQRAE II monitors to use a singleconfiguration.

    Under Operation, select LoadMonitor Configuration.

  • 8/3/2019 Kurzanleitung ARL2 En

    80/91

  • 8/3/2019 Kurzanleitung ARL2 En

    81/91

    Load Monitor ConfigurationLoad Monitor Configuration

  • 8/3/2019 Kurzanleitung ARL2 En

    82/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Always double-check theconfiguration settings to ensurethey are correct.

    If they are not correct, or if duedates have passed, use anotherconfiguration file or create a newone.

    Click Set to load the configuration

    into the AutoRAE Utility.

    Update AutoRAE Lite FirmwareUpdate AutoRAE Lite Firmware

  • 8/3/2019 Kurzanleitung ARL2 En

    83/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    New firmware updates occasionally become available for

    the AutoRAE Lite.

    Download the firmware from the RAE Systems website orfrom a disc to your computer.

    Connect the AutoRAE Lite to the computer using a USB

    cable.

    Update AutoRAE Lite FirmwareUpdate AutoRAE Lite Firmware

  • 8/3/2019 Kurzanleitung ARL2 En

    84/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Power up the AutoRAE Lite (make sure it has batteries or

    has the AC adapter connected and plugged in).

    Make sure a QRAE II is not in the dock. Start the AutoRAE Utility Program by double-clicking its

    icon on the computers desktop.

    Update AutoRAE Lite FirmwareUpdate AutoRAE Lite Firmware

  • 8/3/2019 Kurzanleitung ARL2 En

    85/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Pull down theOperation menu and

    select UpgradeFirmware.

    A pop-up window

    appears, remindingyou to unplug theUSB cable when you

    are done updating thefirmware.

    Click OK.

    Update AutoRAE Lite FirmwareUpdate AutoRAE Lite Firmware

  • 8/3/2019 Kurzanleitung ARL2 En

    86/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    The RAE Programmer 7000 dialog box appears.

    Set the COM port to match your current configuration (for

    example, 6). Click Browse...

    Locate the firmware update file on your computer (it has a

    .rfp suffix).

    Update AutoRAE Lite FirmwareUpdate AutoRAE Lite Firmware

  • 8/3/2019 Kurzanleitung ARL2 En

    87/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Once you select the file, click Open. RAE Programmer 7000 opens.

    Click Go to start updating the firmware.

    You receive the message Processing Information as the updating

    progresses.

    Update AutoRAE Lite FirmwareUpdate AutoRAE Lite Firmware

  • 8/3/2019 Kurzanleitung ARL2 En

    88/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Update AutoRAE Lite FirmwareUpdate AutoRAE Lite Firmware

  • 8/3/2019 Kurzanleitung ARL2 En

    89/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    Update AutoRAE Lite FirmwareUpdate AutoRAE Lite Firmware

  • 8/3/2019 Kurzanleitung ARL2 En

    90/91

    RAE Systems Protection Through DetectionRAE Systems Protection Through Detection

    When the firmware update is complete, you receive the

    message Firmware loaded. Please reset the unit!

    Click Exit.

    Unplug the USB cable, and exit the AutoRAE Utility program.

  • 8/3/2019 Kurzanleitung ARL2 En

    91/91