Ottostraße 20 22 * 63150 Heusenstamm * Germany Phone: +49 (0) 6104 69030 * Fax: +49 (0) 6104 690340 * E-Mail: [email protected] Homepage: www.erweka.com Erweka GmbH; Geschäftsführung: Werner G. Müller, Manfred Koller; Handelsregister: Amtsgericht Offenbach HRB 2382 Instruction Manual Tapped Density Testers SVM 121 / SVM 221 SVM 122 / SVM 222 SVM 223

description

manual de uso y configuracion del tester de densidad

Transcript of SVM121-223 Manual_756-309-0001_Fw1.13_En_V1

Page 1: SVM121-223 Manual_756-309-0001_Fw1.13_En_V1

Ottostraße 20 – 22 * 63150 Heusenstamm * Germany Phone: +49 (0) 6104 69030 * Fax: +49 (0) 6104 690340 * E-Mail: [email protected]

Homepage: www.erweka.com Erweka GmbH; Geschäftsführung: Werner G. Müller, Manfred Koller; Handelsregister: Amtsgericht Offenbach HRB 2382

Instruction Manual Tapped Density Testers

SVM 121 / SVM 221 SVM 122 / SVM 222

SVM 223

Page 2: SVM121-223 Manual_756-309-0001_Fw1.13_En_V1

Ottostraße 20 – 22 * 63150 Heusenstamm * Germany Phone: +49 (0) 6104 69030 * Fax: +49 (0) 6104 690340 * E-Mail: [email protected]

Homepage: www.erweka.com Erweka GmbH; Geschäftsführung: Werner G. Müller, Manfred Koller; Handelsregister: Amtsgericht Offenbach HRB 2382

Document version

Tapped Density Testers SVM 121 / SVM 221 SVM 122 / SVM 222

SVM 223

Date: 2014-01-21 Firmware version: from 1.13

Document No.: 756-309-0001

Document version: 1.0 (English language)

Page 3: SVM121-223 Manual_756-309-0001_Fw1.13_En_V1

Zertifiziertes Managementsystem seit 1996 Managementsystem certified since 1996

ERWEKA GmbH Ottostr. 20-22 D-63150 Heusenstamm Telefon: (06104) 69 03 0 Telefax: (06104) 69 03 40 Geschäftsführer / Managing Director Werner G. Müller Manfred Koller Amtsgericht Offenbach HRB 2382

Page 4: SVM121-223 Manual_756-309-0001_Fw1.13_En_V1

Instruction Manual

Page 4 of 30

Table of Contents

1 INTRODUCTION .............................................................................................................................................. 5

1.1 SYMBOLS USED IN THE MANUAL ..................................................................................................................... 5

2 DELIVERY CHECK .......................................................................................................................................... 6

2.1 LIST OF COMPONENTS .................................................................................................................................... 6

3 ABOUT THE INSTRUMENT ........................................................................................................................... 7

3.1 DESCRIPTION .................................................................................................................................................. 7 3.2 KEYPAD ........................................................................................................................................................... 8 3.3 IDENTIFICATION PLATE .................................................................................................................................... 8 3.4 TECHNICAL DETAILS........................................................................................................................................ 9

4 INSTRUMENT INSTALLATION ................................................................................................................... 10

4.1 SAFETY INSTRUCTIONS................................................................................................................................. 10 4.2 INSTALLATION ............................................................................................................................................... 10 4.3 MAINS SUPPLY .............................................................................................................................................. 11 4.4 SWITCHING THE INSTRUMENT ON ................................................................................................................. 11

5 INSTRUMENT SETUP................................................................................................................................... 12

5.1 THE SND MENU ............................................................................................................................................ 13 5.2 THE LED MENU ............................................................................................................................................ 14 5.3 THE PRNT MENU ......................................................................................................................................... 14 5.4 THE P.TYP MENU ......................................................................................................................................... 16 5.5 THE DATE MENU .......................................................................................................................................... 17 5.6 THE OQ MENU .............................................................................................................................................. 21 5.7 THE VER MENU ............................................................................................................................................ 24

6 PERFORMING THE TESTS ......................................................................................................................... 25

APPENDIX ................................................................................................................................................................ 28

REPORT FROM THE PORTABLE PRINTER ................................................................................................................ 28 REPORT FROM THE CONVENTIONAL PRINTER ........................................................................................................ 29

COPYRIGHT ............................................................................................................................................................. 30

Page 5: SVM121-223 Manual_756-309-0001_Fw1.13_En_V1

SVM 121-223 Instruction Manual Introduction

Page 5 of 30

1 Introduction

Thank you for purchasing the ERWEKA Tapped Density Checker type SVM! To assure continuing and accurate function of the instrument, read this manual thoroughly before operating the instrument and keep it for further reference.

Our products are under continuous development and improvement and we reserve the right to make changes without notification.

Contact ERWEKA at [email protected] for firmware updates, when ordering spare parts, when experiencing technical problems or in the event of the instrument requiring repair. Supply the following information:

type of the instrument

serial number of the instrument

short description of the case (fault)

1.1 Symbols used in the manual

Information: this note provides you with additional information specifically useful at this point of work with the instrument.

Note: this note draws your attention to ensure smooth operation of the instrument.

Page 6: SVM121-223 Manual_756-309-0001_Fw1.13_En_V1

SVM 121-223 Instruction Manual Delivery Check

Page 6 of 30

2 Delivery Check

Upon receiving the instrument, check the completeness of the delivery package prior to installing the instrument. Check the outer package for any physical damage.

If any damage has occurred to the package in transit, inform the manufacturer and the carrier agency IMMEDIATELY.

Unpack the instrument and check it for any evident damage.

If any evident damage to the instrument can be observed, inform the manufacturer IMMEDIATELY.

Check the instrument delivery kit for its completeness. Refer to the list of components below to make sure that none of the components is missing and that all the components have been received in good working condition.

If any of the components or their parts are missing or are damaged, inform the manufacturer IMMEDIATELY.

2.1 List of components

Below is the list of delivered components for the standard SVM:

Name Pieces

Main instrument 1

Measuring cylinder 1 or 2

Support plate 1 or 2

Mains supply cable 1

Instruction manual 1

Page 7: SVM121-223 Manual_756-309-0001_Fw1.13_En_V1

SVM 121-223 Instruction Manual About the Instrument

Page 7 of 30

3 About the Instrument

3.1 Description

The ERWEKA SVM has been designed for measuring the tapped volume and tapped density of powders, granules and similar products. SVM 121 / 221 ERWEKA Tapped Density Testers for the USP method 1 are available with one (SVM 121) and two test stations (SVM 221) for measuring the tapped volume and tapped density of powders, granules and similar products. The instrument features a membrane keypad and two LED displays to enter test parameters. Test parameters can be set in either time (hours/minutes/seconds) or number of strokes.

Only time can be set up for the tapped density testers of SVM 223 type.

The tapped volumeter is equipped with a standard measuring cylinder of 250 ml. Measuring cylinders of 100 ml and 500 ml with suitable adaptors are also available. All electronics employed conforms to EMV guidelines. All electronic wiring fully conforms to VDE 0.100 requirements. SVM 122 / 222 In contrast to the before mentioned SVM 121/221, the SVM 122 (with one test station) and SVM 222 (with two test stations) in respect of strokes/min. and height of stroke meets the USP method 2 and Pharm.Eur. and DIN ISO EN 787/11 standards. SVM 223 ERWEKA Tapped Density Testers, type SVM 223, combines the two different methods in one single unit. The test station 1 operates according to the USP method 1, and the test-station 2 at the same time operates according to the USP method 2/EP/DIN ISO EN 787, part 11.

Page 8: SVM121-223 Manual_756-309-0001_Fw1.13_En_V1

SVM 121-223 Instruction Manual About the Instrument

Page 8 of 30

3.2 Keypad

The keypad of the SVM looks as follows:

The following table outlines the functions and usage of the keys.

Key Symbol Description/Function

To select values; to scroll through menus and submenus

(further referred to as the Stop) To pause a test; to stop a test; to exit the selected menu

(further referred to as the Start) To start a test; to enter the selected menu

All data inputs are supported by a beep tone. The beep tone confirms that the command has been accepted and carried out.

3.3 Identification plate

The identification plate (placed on the back panel of the instrument) of the SVM can look as follows:

The identification label contains the following information:

Type: instrument type

Serial No: instrument number

PN: normal output

IN: rated current

U: voltage

F: frequency

Page 9: SVM121-223 Manual_756-309-0001_Fw1.13_En_V1

SVM 121-223 Instruction Manual About the Instrument

Page 9 of 30

3.4 Technical details

SVM X21 SVM X22 SVM 223

Dimensions: (without measuring cylinder)

Height: [mm] 260

Width: [mm] 300

Depth: [mm] 350

Weight Details:

Weight: [kg] 9.5 10.5 10.5

Voltage:

100 - 240V / 50 – 60 Hz

Safety Fuses:

15A T (slow) 2off 3.15A T (slow) 2off

Testing Stations:

1 2 2

DIN 53194 Requirements:

Working Stroke in mm:

Method: 1 Method: 2

14 2 mm

3 0.2 mm

Speed min-1: Method: 1 Method: 2

300 15

250 15

100 ml measuring cylinder Support plate

130 ± 16 g 240 ± 12 g

250 ml measuring cylinder Support plate

220 ± 44 g 450 ± 10 g

Test Run Timer:

Adjustable up to 9 hours, 59 min, 59 sec.

Stroke Adjustment:

Adjustable up to 9999 No

Control of the Instrument:

Membrane keypad

2 LED displays

Required Installation Environment:

Ambient temperature in operation

+10°C up to +30°C

Storage and transport temperature

-10°C up to +55°C

Relative humidity 25-80% / no condensation

Page 10: SVM121-223 Manual_756-309-0001_Fw1.13_En_V1

SVM 121-223 Instruction Manual Instrument Installation

Page 10 of 30

4 Instrument Installation

4.1 Safety instructions

The instrument is to be plugged in to a safety socket only.

The voltage of the existing current supply is to be compared to the indications on the type label.

ERWEKA accepts no liability in case of wrong connection.

In case of any failure the instrument should be opened by the authorized staff only.

Before opening the instrument, disconnect it from the rear panel.

The SVM instruments produce the noise during their functioning. To avoid disturbance, use a special noise chamber for the instrument or protect your ears.

4.2 Installation

Position the tapped density tester on a horizontal, plane surface, e.g. on a laboratory table.

Remove the support plate from the instrument.

Position the measuring cylinder onto the support plate.

Slide the cap together with the clamp ring over the measuring cylinder, press down slightly and turn to the right until the centering pin locks.

Insert the complete measuring cylinder into the unit and assure that the holding plate locks in.

Assembly drawing

Page 11: SVM121-223 Manual_756-309-0001_Fw1.13_En_V1

SVM 121-223 Instruction Manual Instrument Installation

Page 11 of 30

4.3 Mains supply

The instrument is to be plugged in to a safety mains supply socket only. Compare the voltage of the existing current supply to the indications on the type plate.

For the current supply, connect the enclosed power cord to the socket on the rear side of the instrument and plug this power cord in to the safety socket.

4.4 Switching the instrument on

Turn on the instrument by means of the mains switch on the rear side of the instrument.

Before starting a test, the instrument should be at the correct operational temperature. Switch on the SVM 10 minutes before operating.

In case of any failure, first check the fuse in the mains input socket.

Page 12: SVM121-223 Manual_756-309-0001_Fw1.13_En_V1

SVM 121-223 Instruction Manual Instrument Setup

Page 12 of 30

5 Instrument Setup

All settings for the SVM instrument can be made using the Configuration menu. To access the Configuration menu: 1. Switch the tapped density tester on. 2. The TIME menu is displayed in which you can set up time for your test. Time setting

will be described in the next chapter. Exit the TIME menu by pressing Stop.

3. The CONF (Configuration) menu will appear with its first item, the SND (sound) submenu. To scroll the items of the Configuration menu, use the arrow keys on the right.

4. To enter any of the submenus, select the one you need and press Start. The submenu options will be displayed on the right LED display. To exit the submenu and return to the Configuration menu, press Stop.

The Configuration menu of the SVM instrument contains the following items:

Menu item Function

SND (sound) To switch on/off the beep signal that occurs every time you press the key.

LED (LED contrast) To set the brightness of the LED displays.

PRNT (print settings) To specify how to print the test reports.

DATE (current date) To set the current date.

OQ (operation qualification)

To check that the instrument parameters are as expected (height, number of strokes, run time).

VER (firmware version)

To view the instrument’s firmware version.

EKEY (electronic key)

This menu item is locked.

EXIT To exit the Configuration menu.

Further information about each of these menu items you will find in the following sections.

Page 13: SVM121-223 Manual_756-309-0001_Fw1.13_En_V1

SVM 121-223 Instruction Manual Instrument Setup

Page 13 of 30

5.1 The SND menu

To view/change the sound settings for the instrument:

1. Using the right arrow keys, select the SND item from the Configuration menu.

2. Press Start.

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ 3. Using the right arrow keys, select the needed option and press Start. The SND menu

has the following options:

ON – to enable the beep signals when the data is entered

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

OFF – to disable the beep signals when the data is entered

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

EXIT – to return to the Configuration menu

Page 14: SVM121-223 Manual_756-309-0001_Fw1.13_En_V1

SVM 121-223 Instruction Manual Instrument Setup

Page 14 of 30

5.2 The LED menu

To view/change the LED contrast:

1. Using the right arrow keys, select the LED item from the Configuration menu.

2. Press Start.

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

3. Using the right arrow keys, set the needed LED contrast. The value may range from 0 to 10, and your setting may look as follows on the left.

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ 4. Press Start to confirm the selected value or Stop to exit the LED menu without

changes.

5.3 The PRNT menu

To view/change settings for printing test reports:

1. Using the right arrow keys, select the PRNT item from the Configuration menu.

2. Press Start.

Page 15: SVM121-223 Manual_756-309-0001_Fw1.13_En_V1

SVM 121-223 Instruction Manual Instrument Setup

Page 15 of 30

3. Using the right arrow keys, select the needed option. The PRNT menu includes the following options:

AUTO – to print test reports automatically

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

ASK – to print test reports upon the user confirmation

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

OFF – not to print test reports

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ 4. Press Start to confirm the selected value or select EXIT to return to the

Configuration menu without changes.

Page 16: SVM121-223 Manual_756-309-0001_Fw1.13_En_V1

SVM 121-223 Instruction Manual Instrument Setup

Page 16 of 30

5.4 The P.TYP menu

To select the printer type:

1. Using the right arrow keys, select the P.TYP item from the Configuration menu.

2. Press Start.

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

3. Using the right arrow keys, select the needed option. The P.TYP menu includes the following options:

NORM – conventional printer

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

PORT – small portable printer

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

4. Press Start to confirm the selected value or select EXIT to return to the Configuration menu without changes.

Page 17: SVM121-223 Manual_756-309-0001_Fw1.13_En_V1

SVM 121-223 Instruction Manual Instrument Setup

Page 17 of 30

5.5 The DATE menu

Use this menu to set the actual date and time as well as date format.

The settings in the DATE menu will be saved only if the printing option is available.

To view/change the current date and time:

1. Using the right arrow keys, select the DATE item from the Configuration menu.

2. Press Start.

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

3. The menu includes the following items:

TIME

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

DAY

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

Page 18: SVM121-223 Manual_756-309-0001_Fw1.13_En_V1

SVM 121-223 Instruction Manual Instrument Setup

Page 18 of 30

YEAR

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

FRMT

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

4. Using the right arrow keys, select the TIME menu item to set the actual time. Press Start to enter the menu. The LED displays will show the following on the left.

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

Page 19: SVM121-223 Manual_756-309-0001_Fw1.13_En_V1

SVM 121-223 Instruction Manual Instrument Setup

Page 19 of 30

5. Use the left arrow keys to set the hour value and the right arrow keys to set the minute value. For example, your time setting may look as follows on the left.

6. Press Start to save the specified time

or Stop to return to the DATE menu without changes.

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

7. Using the right arrow keys, select the DAY menu item to set the actual day and month. Press Start to enter the menu. The LED displays will show the following on the left.

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

8. Use the left arrow keys to set the month and the right arrow keys to set the day. For example, your day setting may look as follows on the left.

9. Press Start to save the specified date

or Stop to return to the DATE menu without changes.

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

Page 20: SVM121-223 Manual_756-309-0001_Fw1.13_En_V1

SVM 121-223 Instruction Manual Instrument Setup

Page 20 of 30

10. Using the right arrow keys, select the YEAR menu item to set the current year.

11. Press Start to enter the menu. The

LED displays will show the following on the left.

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

12. Use the right arrow keys to change the default setting with the actual year value.

13. Press Start to save the specified year or Stop to return to the DATE menu without changes.

14. Using the right arrow keys, select the FRMT menu item to set the date format. 15. Press Start to enter the menu. There are the two date formats available:

EUR – dd/mm/yyyy

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

USA – mm/dd/yyyy

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

Page 21: SVM121-223 Manual_756-309-0001_Fw1.13_En_V1

SVM 121-223 Instruction Manual Instrument Setup

Page 21 of 30

16. Use the right arrow keys to set the needed date format. Press Start to save the specified date format or Stop to return to the DATE menu without changes. You can also select EXIT in the FRMT menu to return to the Configuration menu.

5.6 The OQ menu

Use this menu to check the operation qualification (OQ) or, in other words, to verify that the parameters of the instrument are correct. To perform the OQ:

1. Using the right arrow keys, select the OQ item from the Configuration menu.

2. Press Start to enter the OQ menu. The

menu includes the following items:

HGHT (height of stroke)

SPM (strokes/minute)

RTIM (run time)

OQPR (OQ report)

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

3. Using the right arrow keys, select the HGHT menu item to check that the stroke height is appropriate. Press Start.

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

4. The LED displays will show the default height (1.0 mm) as follows on the left. Press Start. The support plate will go up and down.

5. Measure the stroke height with a dial

gauge from the validation tool kit.

Page 22: SVM121-223 Manual_756-309-0001_Fw1.13_En_V1

SVM 121-223 Instruction Manual Instrument Setup

Page 22 of 30

For more information on measuring the stroke height refer to the SVM121-223 Validation Kit Operation Manual.

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ 6. Press Stop to stop the movement of the support plate. The LED display blinks. 7. Enter the actual stroke height shown by the dial gauge and confirm it by pressing

Start. You will be returned to the OQ menu. _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

8. Using the right arrow keys, find the SPM menu item to check that the strokes per minute. Press Start.

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

9. The LED displays will show the default number of strokes (100) as follows on the left.

10. Apply a tachometer from the validation

tool kit to measure the actual number of strokes.

11. Press Start. The support plate will go

up and down. The remaining strokes will be shown in the descending order.

For more information on measuring the number of strokes refer to the SVM121-223 Validation Kit Operation Manual.

12. When the strokes end, the LED display will blink. Enter the actual number of strokes measured by the tachometer and confirm it with Start. You will be returned to the OQ menu.

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

Page 23: SVM121-223 Manual_756-309-0001_Fw1.13_En_V1

SVM 121-223 Instruction Manual Instrument Setup

Page 23 of 30

13. Using the right arrow keys, select the RTIM menu item to check that the instrument’s stop watch works as appropriate and calibrate it, if necessary. Press Start.

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

14. The LED displays will look as follows on the left.

15. Apply a stop watch from the validation

tool kit to measure the time.

16. Press Start to start the time counting in ascending order.

17. Press Stop to stop the time counting.

For more information on measuring the time refer to the SVM121-223 Validation Kit Operation Manual.

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ 18. When the time counting is stopped, the LED display will blink. Enter the actual time

measured by the stop watch and confirm it with Start. You will be returned to the OQ menu.

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

19. If you want to print the report with the OQ test data, using the right arrow keys find the OQPR menu item. Press Start.

To print the report, you should have the printing option available.

Page 24: SVM121-223 Manual_756-309-0001_Fw1.13_En_V1

SVM 121-223 Instruction Manual Instrument Setup

Page 24 of 30

If the printing option is available and enabled but the report is not printed, check that the printer is connected appropriately to the SVM instrument.

20. Press Stop to return to the OQ menu. 21. Using the right arrow keys, select EXIT to return to the Configuration menu.

5.7 The VER menu

Use this menu to view the actual firmware version of the instrument.

The firmware version can be only seen, not changed.

To view the firmware version:

1. Using the right arrow keys, select the VER item from the Configuration menu.

2. Press Start to enter the VER menu.

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

The actual firmware version will be shown on LED display. Press Stop to exit this menu.

Page 25: SVM121-223 Manual_756-309-0001_Fw1.13_En_V1

SVM 121-223 Instruction Manual Performing Tests

Page 25 of 30

6 Performing the Tests

The SVM instruments can perform tests in two modes:

Time (maximum is 9 hours, 59 min, 59 sec.)

Number of strokes (maximum is 9999)

For the tapped density testers of type SVM 223, only the time mode is used.

Before performing the tests, you should do the following preparations:

assemble the instrument following the instructions provided in chapter 4 “Instrument Installation”

fill the measuring cylinder(s) with the product that will be tested Now follow the procedure below. To perform a test: 1. Switch the instrument on.

2. The TIME menu will be shown by default first on the LED displays. If you would like

to perform the test in time mode, use this menu to enter the time required for the test. _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

Using the right arrow keys, enter the needed test time value. At first, the time will be shown in the mm:ss format and each upper arrow key pressing will increase the time value by 1 second. If the value you are entering is more than 59 minutes and 59 seconds, the time will be shown in the hh:mm format and pressing the upper arrow key will increase the displayed time value by 1 minute.

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

Page 26: SVM121-223 Manual_756-309-0001_Fw1.13_En_V1

SVM 121-223 Instruction Manual Performing Tests

Page 26 of 30

If you would like to perform the test in speed mode, when you see the TIME menu, apply the left arrow keys to find the STRK menu for entering the number of strokes required for the test. The LED displays may look as follows on the left. Using the right arrow keys, enter the number of strokes needed for the test.

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

3. After specifying the test duration or strokes, press Start. The test will begin. Depending on the mode you selected, the LED displays may look as follows.

For the test based on time:

The test speed in strokes per minute is shown on the left and the countdown of the set time – on the right.

_ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _

For the test based on the number of strokes:

The test speed in strokes per minute is shown on the left and the countdown of the set number of strokes – on the right.

After the test is finished, the previously set time/stroke value appears on the LED display again. The instrument is ready for a new test. In case the printing option is set to AUTO (see section 5.3 “PRNT menu”), the report with the test results will be printed as soon as the test is finished.

Page 27: SVM121-223 Manual_756-309-0001_Fw1.13_En_V1

SVM 121-223 Instruction Manual Performing Tests

Page 27 of 30

To print the report, you should have the printing option available.

If the printing option is available and enabled, but the report is not printed, check that the printer is connected appropriately to the SVM instrument.

You can interrupt the test by pressing Stop once and end it by pressing Stop twice.

Page 28: SVM121-223 Manual_756-309-0001_Fw1.13_En_V1

SVM 121-223 Instruction Manual Appendix

Page 28 of 30

Appendix

In this chapter you will find the protocol examples printed from the instrument.

Report from the portable printer

Page 29: SVM121-223 Manual_756-309-0001_Fw1.13_En_V1

SVM 121-223 Instruction Manual Appendix

Page 29 of 30

Report from the conventional printer

Page 30: SVM121-223 Manual_756-309-0001_Fw1.13_En_V1

SVM 121-223 Instruction Manual Copyright

Page 30 of 30

Copyright

The material in this document is the intellectual property of ERWEKA. Any copy or reproduction of this document or its parts without the written permission of and reference to ERWEKA is prohibited and will be prosecuted according to the law. Trademarks All trademarks in this manual are the property of their respective owners. Copyright © 2014 by ERWEKA