RTM 2200 Radon/Thoron monitor –” System in a box” for ... · Streil, T , V. Oeser and J....

48
Streil, T , V. Oeser and J. Sabol RTM 2200 Radon/Thoron monitor –” System in a box” for complex sampling procedures and multi parameter analysis SARAD GmbH Wiesbadener Str. 10 D-01159 Dresden [email protected]e

Transcript of RTM 2200 Radon/Thoron monitor –” System in a box” for ... · Streil, T , V. Oeser and J....

Streil, T , V. Oeser and J. Sabol

RTM 2200 Radon/Thoron monitor –” System in a box” for complex sampling procedures and multi

parameter analysis

SARAD GmbHWiesbadener Str. 10

D-01159 [email protected]

2

Selbstorganisierendes Sensor-Netzwerk

1. Personengebundene Sensoren/Detektoren2. Portable Monitore, die bei Bedarf an verschiedenen Orten

platziert werden können

Inhalationsdosisverlauf in der Nähe von Fukushima/Japan

Measuring chamber soil gas probe

Measuring chambers RTM 2100

Radon ScoutRadon Scout PLUS

Radon Scout PMT with Lucas cell

0

2

4

6

8

10

12

14

0

5000

10000

15000

20000

25000

30000

35000

40000

45000

Rad

on c

oncn

etra

tion

[Bq/

m3 ]

test of humidity influence by 20 C

RTM 1688-2 063

RSC PMT 300

theoretic radon decay curve

absolute humidity [g/m3]

Abs

olut

e H

umid

ity [g

/m³]

rel. humidity ~30%

rel. humidity ~ 50 %

rel. humidity ~ 80%

RTM1688-2 Geostation

A full water tight system including soil probe, contain radon/thoron sensor, humidity, temperature, air pressure, tilt

Soil gas probe on aFumarole near Taipei

Measuring chambers of RTM1688-2

Radon / Thoron monitor RTM 1688-2

Outdoor measurement with RTM1688-2 and DOSEMAN ProRadon Thoron

Radon decay products PAEC Thoron decay products PAEC

Intercomparision NIRS average Thoron conc. 4522 Bq/m³ +-102Bq/m3

NIRS_Ref.Rad 7 4428 Bq/m³ +- 217 Bq/m³ NIRS_Ref.AB-5 4610 Bq/m³ +- 575 Bq/m³

DACM System topology

Instrument synthese

DACM-01 Hardware implementation

Component Type Definition

- Definition of the parameter set, later offered to the user(names, units, transfer characteristics, result selection,calibration constants etc.)

- Memory space allocation for basic data- Calculation rules and procedures for results- General result data format

Component properties

Configuration structure

Data structure

- Identification (names, codes etc.)- Possible states of the component type within the cycle (ON/OFF, set-points etc.)- Data and alert handling

Disconnects component functions from their hardware implementation!

Benefits for

Manufacturer:New components can be defined or derived by extensionComponents can be implemented to any hardware platformInstrument specific component sets possible (several types and numbers)Hardware independent configuration and operation software (PC)

User:full control over the whole instrument functionalitysimple re-configuration and extension without manufacturerGeneralised data base for generated results (from simple to very complex)

DACM Dual software concept

Simple configuration interface, administratorknows the structure of the instrument and theintention of the measurement procedure

DACM based instrument looks likea powerful, universal toolbox

Configuration Software

- Cycle Definition- Component Configuration- Calibration work / Testing

DACM based instrument looks likea simple, ready to use instrument

No knowledge about internal functions required

Operation Software

- Module Setup- Cycle Selection- Data Handling- Visualisation/Analysis

OperatorAdministrator

Inter comparison Soil gas measurement Beijing 28.03.2010

0

2000

4000

6000

8000

10000

12000

28/0

3/20

10 …

28/0

3/20

10 …

28/0

3/20

10 …

28/0

3/20

10 …

28/0

3/20

10 …

28/0

3/20

10 …

28/0

3/20

10 …

28/0

3/20

10 …

28/0

3/20

10 …

28/0

3/20

10 …

28/0

3/20

10 …

28/0

3/20

10 …

28/0

3/20

10 …

28/0

3/20

10 …

28/0

3/20

10 …

28/0

3/20

10 …

28/0

3/20

10 …

28/0

3/20

10 …

28/0

3/20

10 …

28/0

3/20

10 …

28/0

3/20

10 …

28/0

3/20

10 …

28/0

3/20

10 …

28/0

3/20

10 …

28/0

3/20

10 …

28/0

3/20

10 …

28/0

3/20

10 …

28/0

3/20

10 …

28/0

3/20

10 …

28/0

3/20

10 …

28/0

3/20

10 …

28/0

3/20

10 …

28/0

3/20

10 …

Rad

on /

Tho

ron

conc

entr

atio

n ( B

q/m

³)

Date / time

Inter comparison Soil gas measurement Beijing 28.03.2010

Thoron RAD 7

Radon RTM 2200

Radon RAD 7

Thoron RTM 2200

Soil gas flux extension with RTM2200

EQF3220 – Radon, Thoron and daughters

Continuous sampling through screen and filter

Removeable sampling head for free positioning inside a room

Three particle size fractions- unattached

Cl t

Quality asurance of the measurement

Screen spectra (unattached daughters)

Cluster (during burning processes)

„Normal“ particle size distribution

PAEC Radon & Thoron daughters

Thoron PAEC Time resolution

Areamonitor A2M4000 –Messung in Gebäude Universiade 2013Kazan

“Hazard-Detection and Management”

7th Dresden SymposiumOrganized by IGRS

Dresden GermanyMarch 3 – March 8, 2013